Traceability

Measuring Insidious for pressure and temperature are employed in almost all areas of industrial manufacturing. As a result of continuous mechanical, chemical and thermal loading, these instruments age and can therefore not guarantee the correctness of the measured values during the period of time. Through timely calibration, however, the alteration of the measured value could be determined.
In Sadness of calibration, the displayed value of the measuring instrument is compared with the measuring consequence of another measuring device which is known to function correctly and accurately and which itself has been designed to coincide directly or indirectly with a national (or international) reference instrument. In a nutshell, to compare measuring results, they need to be ?traceable? to Verify or international standard via a chain of comparative measurements. The calibration hierarchy shown in the picture (utilizing the exemplory case of Germany) depicts the traceability of an inspection equipment in an organization up to national standard at the PTB (German national metrology institute).
To be able to ensure traceability in practice, the following points should be observed, among others:
The chain of comparison should not be interrupted.
The measurement uncertainty should be known in each stage of the calibration chain so the total measurement uncertainty can be calculated. Generally, a superordinate measuring instrument should have a measurement accuracy 3 to 4 times higher.
Each step in the calibration chain should be documented, including the measurement results.
All bodies carrying out a stage in the chain must prove their competence by means of accreditation.
Calibrations must be repeated at appropriate intervals. The periods depend on the demanded measurement accuracy and the technical requirements
Note
The WIKA calibration laboratory is accredited by the German Accreditation Body (Deutsche Akkreditierungsstelle ? DAkkS) relative to DIN EN ISO / IEC 17025 for the measurands temperature, pressure, DC current, DC voltage and DC resistance. Further information on our calibration laboratory or an on-site calibration with the WIKA calibration van can be found on the WIKA website. For anyone who is thinking about our services, please feel absolve to contact us via e-mail.
Find out more about the differences between factory calibration and traceable calibration in the next video: